Is a PID more sensitive than an FID?

A photoionization detector (PID) for gas chromatography (GC), with a proper design should be 50-100 times more sensitive than a flame ionization detector (FID) click on PID-FID ratios.  Some PIDs only have the same or less sensitivity than an FID. The PID Analyzers PID has a dynamic range of 10 exp 8 click on ( 4th-generation-photoionization-detector) and a detection limit of < 1 ppb of benzene   click on (ppt-levels-of-benzene-in-air-by-pid ) .  Contact us at http://www.hnu.com



Is is interesting to note that when we (HNU) introduced the 1st commercial PID in 1976 at #Pittcon in Cleveland, and we stated that  "Our PID is 50 times more sensitive than an FID for benzene", we were greated  with skepticism. Today, 35 years later, many people are still unaware that the PID has this kind of sensitivity.